IC-KP alignment device

Due to the large number of adjustment options, our IC-KP alignment device with adjustment aid enables precise compliance with the specifications that are prescribed for the various prosthesis components.
It is made of stainless steel and has a shut-off bladder for shaft fixation. The knee joint can also be fixed and a footrest is available. The integrated barometer measures the pressure in the shut-off bladder (max. 1.5 bar).
There are also 2 linear rails, one horizontal and one vertical. The ball joints are used for abduction, adduction, flexion and extension. Measuring tools, heel wedges, etc. can be stored in the drawer under the footrest.
The alignment device should be fixed locally, preferably on a wall or workbench.
Use
The construction and the adjustments may only be done by an orthopedic technician, as he has the necessary specialist knowledge and can therefore produce an optimal prosthesis.
Maintenance
If you maintain and clean it yourself, the device is maintenance-free. Please note that the pressure reducer must not exceed 1.5 bar. In addition, the shut-off bladder should be checked regularly for wear.
With the help of a laser plummet, you can now position the prosthesis components in relation to one another in such a way that they meet the construction criteria of the components and the individuality of the patient.